domains / families / science / radar-vital-signs
End-to-end pipeline for extracting heart rate (HR) and breathing rate (BR) from raw short-range radar I/Q captures — both continuous-wave (CW, 24 GHz clinical boards) and FMCW mmWave (60/77 GHz, TI IWR/AWR). Use when Claude needs to parse interleaved I/Q binary, do a Range FFT on FMCW chirps, remove static clutter, pick a subject range bin, extract phase with unwrapping, design HR/BR bandpass filters, pick a peak frequency via PSD, reject the HR second harmonic that often dominates the fundamental, or handle respiration-harmonic leakage on slow breathers. Not for pulse/UWB range gating, MIMO beamforming, Doppler-only gesture radar, arrhythmia detection, or multi-subject source separation.
| Repository | Path |
|---|---|
| benchflow-ai/skillsbench | tasks/radar-vital-signs/environment/skills/radar-vital-signs/SKILL.md |